Construction and application of service quality evaluation system in the preclinical research on cardiovascular implant devices

Abstract

Background

Services for the preclinical development and evaluation of cardiovascular implant devices (CVIDs) is a new industry. However, there is still no indicator system for quality evaluation. Our aim is to construct a service for quality evaluation system for the preclinical research and development of CVIDs based on Fuzzy Analytical Hierarchy Process (FAHP).

Methods

First, we reviewed the related literature to identify and select possible factors. Second, we developed an analytic hierarchy process framework. Third, we developed a questionnaire based on pairwise comparisons and invited 10 experienced specialists to rate these factors. We then used FAHP to compute the weights of these factors and prioritize them. Finally, to demonstrate the effectiveness of the proposed indicator system, a case study was performed as a practical example.

Results

Four main indicators (professionalism, functionality, stability and security) and 15 subindicators were selected to form the service evaluation system based on literature review and expert’s proposals. According to the weight calculation data, the order of primary indicators by importance, is professionalism (0.6457), security (0.1193), functionality (0.0958) and stability (0.0596) in sequence. Top five secondary indices are personnel’s technical ability, facility and equipment attractiveness, data auditability, confidentiality capability and professional service procedures. In the case study, FW’s final actual effectiveness value was 0.9076, which is the same as the actual situation.

Conclusion

The indicator system established in this study is comprehensive, reasonable, reliable and with strong practicality. It is worth popularizing and applying. The implementation of this evaluation system can provide measurable evidence for service demander and a way to improve service quality for suppliers.

Peer Review reports

Background

Cardiovascular disease is the leading cause of death and disability worldwide [1,2,3,4]. Implantation of high-quality medical devices is one of the most commonly used means for the treatment of cardiovascular diseases. With the increasing morbidity and mortality of cardiovascular disease, there is an increasing market demand for high-quality cardiovascular implant devices (CVIDs), such as coronary artery stent, heart valves and artificial assistant equipment [5,6,7,8].

CVIDs belong to high-risk class III medical devices, which are characterized by security primacy, technical complexity, professional operation, long development line and high cost. Therefore, the need for a new service industry for the preclinical development and evaluation of implant devices arises at this historic moment. However, there is still no service quality evaluation system, which might be the first step to the smooth, efficient running and standardization of this new industry [9].

According to previous studies, Analytic Hierarchy Process is a useful method for analysis of factors influencing service [10], and medical decision support [11,12,13,14]. The fuzzy theory has also been recommended for its ability to prevent expert judgment from being influenced by extreme values, to combine the participants’ opinions more reasonably, as well as both objectively and accurately to prioritize the relevant indicators and to calculate their weight values under a hierarchy model [15,16,17,18]. This suggests to us that the Fuzzy Analytic Hierarchy Process (FAHP) has the advantages of both AHP and fuzzy theory, and it maybe an effective method for establishing a service quality evaluation system for the preclinical development of CVIDs [19,20,21].

In the present study, we will focus on establishing a service quality evaluation system for the preclinical development of cardiovascular implant devices, based on FAHP. This would contribute to strengthening quality control and regulating the behavior in preclinical research and development of medical device, so as to guarantee that the public can safely and effectively use them [22, 23].

Methods

Arrangement of a decision-making group (DMG)

First, a DMG was organized to outline a structure for the process indicators according to the previously reported inclusive criteria [24, 25] as follows:

  1. 1)

    Having an academic degree (PhD, MSc) in one of the mentioned majors: clinical, management.

  2. 2)

    Working as faculty in a hospital or Research Institute, associated with cardiovascular implant devices.

  3. 3)

    Having empirical studies in subjects relevant to cardiovascular device development and application.

  4. 4)

    willing to answer the expert consultation form

Establishment of the evaluation framework

Then, we comprehensively reviewed literature [26,27,28,29,30,31] and collected the comments of the DMG. Indicators that could most effectively reflect the service quality in the preclinical development of cardiovascular implant devices were selected for constructing the two- index-hierarchy indicator system. The primary dimensions included professionalism, security, functionality and stability. The secondary dimensions included personnel’s technical ability, hardware attractiveness, professional service procedures, permission suitability, confidentiality capability of information and resources, etc.

Calculation of the weights and ranks of the indicators

Construct pair-wise comparison (PWC) matrix

With regard to the service validity evaluation framework, the DMG were required to perform pair-wise comparisons between the first and second levels of indicators, on the basis of their knowledge. To do this, they compared the importance of each indicator of both the first and second levels, with the adjacent indicators of its own level. Then, a matrixX was created according to the pair-wise comparisons [32,33,34].

$$ {\tilde{x}}_{ij}=\left[\begin{array}{cccc}{\tilde{x}}_{11}& {\tilde{x}}_{12}& \cdots & {\tilde{x}}_{1n}\\ {}{\tilde{x}}_{21}& {\tilde{x}}_{22}& \cdots & {\tilde{x}}_{2n}\\ {}\vdots & \vdots & \ddots & \vdots \\ {}{\tilde{x}}_{n1}& {\tilde{x}}_{n2}& \cdots & {\tilde{x}}_{nn}\end{array}\right] $$
(1)

Comparison matrix consistency check

After the comparison, matrices were established, and the consistency checks of the matrices was performed by computing the consistency ratio (CR):

$$ \mathrm{CR}=\frac{\mathrm{CI}}{\mathrm{RI}};\mathrm{CI}=\left({\uplambda}_{max}-n\right)/\left(n-1\right) $$
(2)

Where: λmax is the largest Eigen value of the comparison matrix. “CI” indicates the consistency index,“RI” denotes the random index, and “n” is the number of criteria that would be judged against (i.e., matrix size).

Determination of the indicators’ weights

Linguistic variables are used in the questionnaire to convert the measured qualitative factors to fuzzy numbers (see Additional file 1). The linguistic variables chosen are commonly used variables - equally important (EI), weakly more important (WMI), strongly more important (SMI), very strongly more important (VSMI), and absolutely more important (AMI). To score the importance of indicators influencing the service quality for preclinical research on cardiovascular implant devices, 1/9–9 scaling method was used as the scoring principle which showed the relative importance of the former indicator (A) compared with the latter indicator (B). All of the primary indicators and the secondary indicators were paired and compared respectively. The data were shown in Additional file 2.

The pair-wise comparison matrix between criteria is then formed based on the fuzzy numbers to evaluate the weights using the FAHP method [35,36,37,38,39]. Based on the previously constructed pair-wise comparison matrixxij, the weights determined are as follows [24]:

$$ {\alpha}_j{\left[\underset{j=1}{\overset{n}{\Pi}}{l}_{ij}\right]}^{1/n};{\beta}_j={\left[\underset{j=1}{\overset{n}{\Pi}}{m}_{ij}\right]}^{1/n};{\gamma}_j{\left[\underset{j=1}{\overset{n}{\Pi}}{n}_{ij}\right]}^{1/n};{\delta}_j={\left[\underset{j=1}{\overset{n}{\Pi}}{s}_{ij}\right]}^{1/n} $$
(3)

and

$$ \alpha =\underset{j=1}{\overset{n}{\Sigma}}{\alpha}_j;\beta =\underset{j=1}{\overset{n}{\Sigma}}{\beta}_j;\gamma =\underset{j=1}{\overset{n}{\Sigma}}{\gamma}_j;\delta =\underset{j=1}{\overset{n}{\Sigma}}{\delta}_j $$
(4)

We then prioritized the extracted indicators of service quality in a hierarchy model identified by the FAHP approach [4] (see Additional file 3).

Results

Demographic characteristics of experts

A total of 10 experts with a senior professional title in the internal (n = 5) and surgery (n = 5) departments of national center for cardiovascular diseases, were invited to participate in this study. Half male and half female, most belonged to the 30–50 age group. Six expert’s work experience was 5–10 years (60%), 3 expert’s work experience was 10–20 years (30%) and 1 expert’s work was more than 20 years (10%). Information describing the DMs is presented in Table 1.

Table 1 Descriptive demographic characteristics of specialists

Extracting the affecting dimensions of service quality

After a comprehensive review of the literature [17, 19, 21] and consideration of the DMG’s opinions, we selected indicators that can be applied to effectively assess the service quality in preclinical development of cardiovascular implant devices. Thus a two-level evaluation system, including primary and secondary dimensions, was established.

As shown in Table 2 and Fig. 1, the primary dimensions include professionalism, security, functionality and stability. With respect to the DMG expertise, we further developed some secondary indicators for each primary dimension. “Professionalism” indicators include personnel’s technical ability, hardware attractiveness, professional service procedures and brand image. Security includes the permission suitability, readiness, auditability and confidentiality capability of information and resources. Functionality includes functional integrity, sufficiency, reasonable interactive communication mechanism and applicance. Stability includes service continuity, stability, and report timely submission rate.

Table 2 Hierarchical structure of indicators and sub-indicators
Fig. 1
figure 1

The two-level evaluation system. Based on literature review and DMG’s opinions, the two-level evaluation system was established. The first level indicator includes 4 items, and the second level indicator includes 15 items. DMG means decision-making group

Extracting the weights and important coefficients of service quality indicators

The weight of each evaluation indicator was calculated. Accordingly, the order of primary indicators, from top to bottom, is professionalism (0.6457), security (0.1193), functionality (0.0958) and stability (0.0596). The top five sub-dimemsions, in sequence, are personnel’s technical ability, hardware attractiveness, data auditability, confidentiality capability, professional service procedures and project compliance (Table 3).

Table 3 Weight and prioritization of indicators and sub-indicators using FAHP

In this paper, to verify the consistency and validity of expert scoring results, the consistency ratio (CR) was used. CR enables observation of variations between the different pair-wise comparisons. In general, the CR value of the pairwise comparison matrix being less than or equal to 0.1 indicates that the expert’s judgments are reasonable; above 0.1 means weak consistency [40]. The results of this study show that the consistency of expert responses is 0.01, which means that the confidence level is over 90%. It is thus concluded that the responses expressed by the experts are well thought out rather than subjectively determined.

Service quality grading

For the convenience application, a service quality grading method was developed in this study. First, the experts graded the formula calculation results (FCR) according to the formulas in Table 4. Second, the actual effectiveness value (AEV) of each indicator was achieved by multiplying the FCR times their own weights. Finally, laboratory service ability was graded according to the standards expressed using “star level”. A total of five service quality levels were suggested, in which level 5 was the highest and level 1 was the lowest in the hierarchy (see Table 5 for details). If needed, the level evaluation method can not only evaluate primary indicators, but also independently evaluate secondary indicators.

Table 4 Calculation of different index items for “star” grading
Table 5 the presentation of star rating results

Case study

To verify the practical applicability of the proposed service quality evaluation system, a case study was performed. A questionnaire was designed to collect the DMG’s judgments. The DMG’s were then asked to evaluate the service supplier, FW lab, according to the requirements in Table 6.

Table 6 Comprehensive evaluation for pre-clinical service quality evaluation of cardiovascular implantation in FW

Given that the FW lab undertakes projects from both clinical researchers and domestic/abroad enterprises, who engage in the research & development of cardiovascular implants devices. Their customers mainly include clinical doctors (project manager), graduates (project executor), and R&D staff of companies. Therefore, in this case study, respondents included different types of customers: clinical doctors (n = 10), graduates (n = 10), the company research and development staff (n = 10). A total of 30 questionnaires were handed out, and the recovery number of valid questionnaires was 30 copies. The average age was 35 years old, the sex ratio 3:1 (male: female), all had obtained either a bachelor’s degree, graduate degree or above, and accounted for 75%. According to the survey data and interview contents, the FCR of each index was calculated; combined with primary indicators and sub-indicator’s weight, the comprehensive implementation effectiveness assessment set was obtained. After normalization processing and further assignment, FW’s final comprehensive service quality AEV was 0.9076, therefore, we concluded that the comprehensive service ability of FW for the preclinical development of cardiovascular implant devices is up to five-star level.

Discussion

Quality is regarded as an important factor in all organizations especially those involving patient life and health. As for the newly arising industry of services for the preclinical development and evaluation of implant devices, there is still no quality evaluation system. In this study, we used the FAHP method to construct a service quality evaluation indicator system, which will provide a method for the service demanders to select ideal suppliers, and for the service suppliers to improve their service quality.

The adopted FAHP technique in this study is one of the most widely used multi-criteria decision making methods [10]. It has been proposed for medical diagnosis, evaluation and selection of medical treatments and therapies; however, no studies have been done with the service quality evaluation for the preclinical development of cardiovascular implant devices. The conventional AHP only takes into account the distinct judgments of decision makers, [18] but it can’t fully reveal human’s fuzzy opinions [16]. So, FAHP method, a fuzzy extension of AHP, was developed by integrating fuzzy comparison ratios. The fuzzy set theory, puts together the comparison process more flexibly and potently in order to clarify experts’ preferences [20]. Practice has proven that it is equally suitable for building a feasible and reasonable service quality evaluation indicator system in our study.

In this study, both customer and service supplier factors were comprehensively considered [41] and a relatively complete indicator system was constructed through FAHP analysis. The indicator system consists of two index hierarchies. Primary dimensions include professionalism, security, functionality and stability. Among them, professionalism is the most important with the highest weight values in the primary indicators, based on our research findings.

Professionalism can be reflected in four aspects: brand image, personnel’s technical ability, facility and equipment attractiveness, and professional service procedures. Brand image refers to the personality characteristics of the company or a certain brand in the market and in the public’s heart. It best reflects the supplier’s professionalism and the public’s recognition of the brand, especially the consumers [42]. However, personnel’s technical ability and facility/equipment attractiveness are the foundation of professionalism [43]. To enhance the brand image and service professionalism, it is an indispensably important content and method to improve personnel’s technical ability and hardwares. With increasingly fierce market competition, professional service procedures have become another core of corporate competition and have become an important strategy for the entire brand [44]. A good sense of service (reputation) can win more customers for the company, which is bound to enhance the market competitiveness of the company.

The top five secondary indicators in the service quality evaluation system include personnel’s technical ability, facility and equipment attractiveness, data auditability, confidentiality capability and professional service procedure in order. Among which, personnel’s technical ability, facility and equipment attractiveness and professional service procedure belong to professionalism as mentioned above. Data auditability and confidentiality capability belong to the primary dimension “security”. Auditability requires a variety of records, which are an important part of the traceability. As for confidentiality, service supplier should establish a secure strategy and system, and all of the personnel should understand the secure strategy and system requirements. This is an important indicator proposed from the perspective of customer requirement.

For the convenience of popularization and application of this system, this study also proposed the use of a “star” system to represent service quality evaluation results. As a case study, this study used the system to assess the service quality of FW lab which is the largest cardiovascular implanted devices preclinical research and development service laboratory in China. The results show that the evaluation results obtained by this system are consistent with the actual survey results.

In terms of enterprise brand image, most of the domestic cardiovascular implant device manufacturers are cooperating with the FW laboratory. FW laboratory has obtained the highest level of qualification, and occupied more than 60% market share. In terms of professional service personnel, the laboratory has a total of 40 employees, all of whom are qualified. More than half of the people have obtained a professional bachelors degree or above, with a solid basis and rich practical experience in team work. FW Laboratory has the first domestic one-stop hybrid operating room dedicated to animal experiment (cleanliness up to grade 10.000). It is equipped with 3 mobile C arm X-ray machines (American GE), the Dutch Philips real-time three-dimensional ultrasound machine, Germany Drager anesthesia machine, a breathing machine and other advanced equipment, which are attractive enough.

Based on the long-term service practice, FW laboratory has accumulated a wealth of experience, mastered advanced technologies and established professional service procedure. The integrity and stability of service, experimental data availability and readiness are satisfactory, which were widely recognized. It is really up to five-star level laboratory. According to the evaluation results, a proposal for improving its service was put forward: because of high frequency use, some precise instruments should be checked regularly to avoid impact on service continuity.

This study has several limitations. Our experts were limited to Beijing shanghai and Tianjin. Future research can be in a more culturally diverse geographical region and compared with the results of this paper since preferences/experiences may change by country, tradition or socioeconomic level. More studies are required to investigate the applicability of the indicators of service quality evaluation system developed in this paper.

Conclusions

The service quality evaluation system constructed in this study is effective and can be popularized. Application of this system will provide a measurable basis for the service demander to select service supplier and provide a method for the supplier to improve their service quality.

Abbreviations

AMI:

Absolutely more important

CR:

Consistency ratio

CVIDs:

Cardiovascular implant devices

DMG:

Decision-making group

EI:

Qually important

FAHP:

Fuzzy Analytical Hierarchy Process

PWC:

Pair-wise comparison

SMI:

Strongly more important

VSMI:

Very strongly more important

WMI:

Weakly more important

References

  1. Santulli G. Epidemiology of cardiovascular disease in the 21 st century: updated numbers and updated facts. J Cardiovasc Dis. 2013;1(1):1–2.

    Google Scholar 

  2. Wilkins E, Wilson L, Wickramasinghe K, Bhatnagar P, Leal J, Luengo-Fernandez R. Townsend, N. European Cardiovascular Disease Statistics 2017 edition. Brussels: European Heart Network; 2017. p. 192. https://www.statista.com/study/44317/european-cardiovascular-disease-statistics-2017/.

  3. Mozaffarian D, Benjamin EJ, Go AS, Arnett DK, Blaha MJ, Cushman M, Turner MB. Heart disease and stroke statistics-2015 update: a report from the American Heart Association. Circulation. 2015;131(4):e29–39. https://doi.org/10.1161/CIR.0000000000000152.

    PubMed  Google Scholar 

  4. Li H, Ge J. Cardiovascular diseases in China: current status and future perspectives. IJC Heart Vasc. 2015;6:25–31. https://doi.org/10.1016/j.ijcha.2014.10.002.

    Article  Google Scholar 

  5. Grand View Research, Angioplasty Balloons Market Size, Share & Trends Analysis Report By Type, By Application, By End Use, And Segment Forecasts, 2018–2025. 2018;25–30 Report ID: GVR-1-68038-356-0. https://www.grandviewresearch.com/industry-analysis/angioplasty-balloons-market.

  6. Grand View Research Connected Drug Delivery Devices Market Analysis Report By Technology, By Product, By End User, And Segment Forecasts, 2018–2025. 2018, 46–50. Report ID: GVR-2-68038-651-6. https://www.grandviewresearch.com/industry-analysis/connected-drug-delivery-devices-market.

  7. Grand View Research Minimally Invasive Surgical Instruments Market Size, Share & Trends Analysis Report by Device, Application, By End Use, And Segment Forecasts, 2018 - 2025, 2018.13–18. Report ID: GVR-2-68038-609-7. https://www.grandviewresearch.com/industry-analysis/minimally-invasive-surgical-instruments-market.

  8. Grand View Research, Hi-Tech Medical Devices Market Analysis Report By Product, By Site, And Segment Forecasts, 2018–2025. 2018.13–16. Report ID: GVR-2-68038-461-1. https://www.grandviewresearch.com/industry-analysis/hi-tech-medical-devices-market.

  9. IMDRF. International Medical Device Regulators Forum. IMDRF Website, 2017. (August), 1–17. Retrieved from http://www.imdrf.org/index.asp.

  10. Shafii M, Rafiei S, Abooee F, Bahrami MA, Nouhi M, Lotfi F, Khanjankhani K. Assessment of service quality in teaching hospitals of Yazd University of Medical Sciences: using multi-criteria decision making techniques. Osong Public Health and Research Perspectives. 2016;7(4):239–47. https://doi.org/10.1016/j.phrp.2016.05.001.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Sloane EB, Liberatore MJ, Nydick RL. Medical decision support using the Analytic Hierarchy Process. Journal of Healthcare Information Management: JHIM. 2002;16(4):38–43. Retrieved from http://www.ncbi.nlm.nih.gov/pubmed/12365298.

    PubMed  Google Scholar 

  12. Schmidt K, Aumann I, Hollander I, Damm K, Von Der Schulenburg JMG. Applying the analytic hierarchy process in healthcare research: a systematic literature review and evaluation of reporting. BMC Medical Informatics and Decision Making. 2015. https://doi.org/10.1186/s12911-015-0234-7.

  13. Padilla-Garrido N, Aguado-Correa F, Cortijo-Gallego V, López-Camacho F. Multicriteria decision making in health care using the analytic hierarchy process and microsoft excel. Med Decis Mak. 2014;34(7):931–5. https://doi.org/10.1177/0272989X14534533.

    Article  Google Scholar 

  14. Hancerliogullari G, Hancerliogullari KO, Koksalmis E. The use of multi-criteria decision making models in evaluating anesthesia method options in circumcision surgery. BMC Medical Informatics and Decision Making. 2017;17(1):1–13. https://doi.org/10.1186/s12911-017-0409-5.

    Article  Google Scholar 

  15. Goepel KD. Implementing the analytic hierarchy process as a standard method for multi-criteria decision making in corporate enterprises – a new AHP excel template with multiple inputs. Proceedings of the International Symposium on the Analytic Hierarchy Process, 2013 of Medical Devices. 2005. p. 1–10. Retrieved from https://www.researchgate.net/publication/275584446_Implementing_the_Analytic_Hierarchy_Process_as_a_Standard_Method_for_Multi-Criteria_Decision_Making_in_Corporate_Enterprises_-_a_New_AHP_Excel_Template_with_Multiple_Inputs. https://doi.org/10.1586/17434440.2.1.41.

  16. Chang CC, Liang C, Shu KM, Tsai CW. Key successful factors of knowledge management for university students using e-portfolios: approach of fuzzy Delphi and fuzzy AHP. Comput Appl Eng Educ. 2015;23(5):673–681. https://doi.org/10.1002/cae.21640.

  17. Chen M-K, Wang S-C. The use of a hybrid fuzzy-Delphi-AHP approach to develop global business intelligence for information service firms. Expert Syst Appl. 2010;37(11):7394–407. https://doi.org/10.1016/j.eswa.2010.04.033.

    Article  Google Scholar 

  18. Ma Z, Shao C, Ma S, Ye Z. Constructing road safety performance indicators using fuzzy delphi method and grey delphi method. Expert Syst Appl. 2011;38(3):1509–14. https://doi.org/10.1016/j.eswa.2010.07.062.

    Article  Google Scholar 

  19. Boyle J, Maeder A, Bell J, John S, Scott R. Medical device research. Expert Review of Medical Devices. 2005. https://doi.org/10.1586/17434440.2.1.41.

  20. Parasuraman A, Zeithaml VA, Berry LL. A conceptual model of service quality and its implications for future research. J Mark. 1985;49(4):41. https://doi.org/10.2307/1251430.

    Article  Google Scholar 

  21. Sharples S, Martin J, Lang A, Craven M, O’Neill S, Barnett J. Medical device design in context: a model of user-device interaction and consequences. Displays. 2012;33(4–5):221–32. https://doi.org/10.1016/j.displa.2011.12.001.

    Article  Google Scholar 

  22. Kim CE, Shin JS, Lee J, Lee YJ, Kim MR, Choi A, Park KB, Lee HJ, Ha IH. Quality of medical service, patient satisfaction and loyalty with a focus on interpersonal-based medical service encounters and treatment effectiveness: a cross-sectional multicenter study of complementary and alternative medicine (CAM) hospitals. BMC Complement Altern Med. 2017;17:174. https://doi.org/10.1186/s12906-017-1691-6.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Koenig SC, Jimenez JH, West SD, Sobieski MA, Choi Y, Monreal G, Slaughter MS. Early feasibility testing and engineering development of a sutureless beating heart connector for left ventricular assist devices. ASAIO J. 2014;60(6):617–25. https://doi.org/10.1097/MAT.0000000000000134.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Cronin JJ, Taylor SA. SERVPERF versus SERVQUAL: reconciling performance-based and perceptions-minus-expectations measurement of service quality. J Mark. 1994;58(1):125. https://doi.org/10.2307/1252256.

    Article  Google Scholar 

  25. Kang G-D, James J. Service quality dimensions: an examination of Grönroos’s service quality model. Manag Serv Qual. 2004;14(4):266–77. https://doi.org/10.1108/09604520410546806.

    Article  Google Scholar 

  26. Altenstetter C. Medical device regulation in the European Union, Japan and the United States. Commonalities, differences and challenges. Innovation: The European Journal of Social Science Research. 2012;25(4):362–88. https://doi.org/10.1080/13511610.2012.723328.

    Google Scholar 

  27. Cronin EM, Varma N. Remote monitoring of cardiovascular implanted electronic devices: a paradigm shift for the 21st century. Expert Review of Medical Devices. 2012. https://doi.org/10.1586/erd.12.18.

  28. Gustafsson F, Rogers JG. Left ventricular assist device therapy in advanced heart failure: patient selection and outcomes. Eur J Heart Fail. 2017. https://doi.org/10.1002/ejhf.779.

  29. Gregory SD, Timms D, Gaddum N, Mason DG, Fraser JF. Biventricular assist devices: a technical review. Ann Biomed Eng. 2011. https://doi.org/10.1007/s10439-011-0348-8.

  30. Kramer DB, Xu S, Kesselheim AS. Regulation of medical devices in the United States and European Union. N Engl J Med. 2012;366(9):848–55. https://doi.org/10.1056/NEJMhle1113918.

    Article  CAS  PubMed  Google Scholar 

  31. Lee S, Fukamachi K, Golding L, Moazami N, Starling RC. Left ventricular assist devices: from the bench to the clinic. Cardiology. 2013;125(1):1–12. https://doi.org/10.1159/000346865.

    Article  PubMed  Google Scholar 

  32. Hung YH, Huang ML, Chen KS. Service quality evaluation by service quality performance matrix. Total Qual Manag Bus Excell. 2003;14(1):79–89. https://doi.org/10.1080/14783360309706.

    Article  Google Scholar 

  33. Chen L-F, Chen S-C, Su C-T. An innovative service quality evaluation and improvement model. Serv Ind J. 2018;38(3–4):228–49. https://doi.org/10.1080/02642069.2017.1389907.

    Article  Google Scholar 

  34. Dagger TS, Sweeney JC, Johnson LW. A hierarchical model of health service quality: scale development and investigation of an integrated model. J Serv Res. 2007;10(2):123–42. https://doi.org/10.1177/1094670507309594.

    Article  Google Scholar 

  35. Lai IKW, Hitchcock M. A consideration of normality in importance–performance analysis. Curr Issue Tour. 2015;18(10):979–1000. https://doi.org/10.1080/13683500.2014.990423.

    Article  Google Scholar 

  36. Azzopardi E, Nash R. A critical evaluation of importance-performance analysis. Tour Manag. 2013;35:222–33. https://doi.org/10.1016/j.tourman.2012.07.007.

    Article  Google Scholar 

  37. Feng M, Mangan J, Wong C, Xu M, Lalwani C. Investigating the different approaches to importance-performance analysis. Serv Ind J. 2014. https://doi.org/10.1080/02642069.2014.915949.

  38. Sever I. Importance-performance analysis: a valid management tool? Tour Manag. 2015;48:43–53. https://doi.org/10.1016/j.tourman.2014.10.022.

    Article  Google Scholar 

  39. Yu KT, Shen CY. Service quality management by expanded traditional importance-performance analysis. Journal of the Chinese Institute of Industrial Engineers. 2011;28(2):146–54. https://doi.org/10.1080/10170669.2010.546557.

    Article  Google Scholar 

  40. Saaty TL. The analytic hierarchy process: planning, priority setting, resources allocation. McGraw-Hill Inc.; 1980. Retrieved from http://linkinghub.elsevier.com/retrieve/pii/0377221782900224.

  41. Shen CY. Strategic vendor selection criteria discussed in relation to demand and supply perspectives. Journal of Industrial and Production Engineering. 2014;31(7):405–16. https://doi.org/10.1080/21681015.2014.975161.

    Article  Google Scholar 

  42. Ramseook-munhurrun P, Lukea-bhiwajee SD, Naidoo P. Service quality in the public service. International Journal of Marketing and Marketing Research. 2010;3(1):37–50. https://doi.org/10.1017/CBO9781107415324.004.

    Google Scholar 

  43. Völckner F, Sattler H, Hennig-Thurau T, Ringle CM. The role of parent brand quality for service brand extension success. J Serv Res. 2010;13(4):379–96. https://doi.org/10.1177/1094670510370054.

    Article  Google Scholar 

  44. Gorla N, Somers TM, Wong B. Organizational impact of system quality, information quality, and service quality. J Strateg Inf Syst. 2010;19(3):207–28. https://doi.org/10.1016/j.jsis.2010.05.001.

Download references

Acknowledgments

None.

Funding

This work was supported by Beijing Municipal Science & Technology Commission [grant numbers Z161100005016014, Z101107052210004], Beijing Key Laboratory of Preclinical Research and Evaluation of Cardiovascular Implant Materials (grant number: 2018-PT2-ZR04) and China Union Medical College Small-scale Characteristic School Project (grant number: 2017-E-JG02). The funding body did not involve in the design of the study or collection, analysis, and interpretation of data.

Availability of data and materials

All data generated or analysed during this study are included in this published article [and its supplementary information files].

Author information

Authors and Affiliations

Authors

Contributions

YCC, YT, XW and JJL have contributed in study design. FLL, BQ Y and BL have contributed in acquisition and analysis of data. GXY has contributed in interpretation of data. YCC and QZ have contributed in drafting and development of the final manuscript. Gd has critically evaluated the draft article. All authors reviewed and approved final version of the manuscript.

Corresponding authors

Correspondence to Yongchun Cui, Yue Tang or Xin Wang.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional files

Additional file 1:

Expert score sheet. To perform pair-wise comparisons between the first and second levels of indicators, linguistic variables were used in the sheet to convert the measured qualitative factors to fuzzy numbers. EI: equally important; WMI: weakly more important; SMI: strongly more important; VSMI: very strongly more important; AMI; absolutely more important. (PDF 392 kb)

Additional file 2:

Raw data for expert scoring. A total of 10 experts were invited to participate in this study. The score sheet recovery rate was 100%. The numbers in the last column represented the average of expert scores. (PDF 85 kb)

Additional file 3:

Raw data for the weight and priority of indicators and subindicators using FAHP. With respect to the expert scoring results, the weight and priority of each evaluation indicator were calculated. The order of primary indicators, from top to bottom, was professionalism, security, functionality and stability. The top five sub-dimemsions included personnel’s technical ability, hardware attractiveness, data auditability, confidentiality capability, professional service procedures and project compliance. (PDF 56 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cui, Y., Luo, F., Yang, B. et al. Construction and application of service quality evaluation system in the preclinical research on cardiovascular implant devices. BMC Med Inform Decis Mak 19, 37 (2019). https://doi.org/10.1186/s12911-019-0773-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12911-019-0773-4

Keywords