Improved delineation model of a standard 12-lead electrocardiogram based on a deep learning algorithm

Abstract

Background

Signal delineation of a standard 12-lead electrocardiogram (ECG) is a decisive step for retrieving complete information and extracting signal characteristics for each lead in cardiology clinical practice. However, it is arduous to manually assess the leads, as a variety of signal morphological variations in each lead have potential defects in recording, noise, or irregular heart rhythm/beat.

Method

A computer-aided deep-learning algorithm is considered a state-of-the-art delineation model to classify ECG waveform and boundary in terms of the P-wave, QRS-complex, and T-wave and indicated the satisfactory result. This study implemented convolution layers as a part of convolutional neural networks for automated feature extraction and bidirectional long short-term memory as a classifier. For beat segmentation, we have experimented beat-based and patient-based approach.

Results

The empirical results using both beat segmentation approaches, with a total of 14,588 beats were showed that our proposed model performed excellently well. All performance metrics above 95% and 93%, for beat-based and patient-based segmentation, respectively.

Conclusions

This is a significant step towards the clinical pertinency of automated 12-lead ECG delineation using deep learning.

Peer Review reports

Introduction

The assessment of electrocardiogram (ECG) signals’ waveform morphology is a crucial step designed to assist cardiologists in diagnosing heart diseases [1]. The diverse morphology of heart diseases is becoming more complicated, making the construction of an automated delineation algorithm challenging [2, 3]. ECG delineation plays a vital role in providing amplitudes, ranges, duration, and morphology [4], and it aims to determine the location of the peaks and boundaries (onset and offset) of three main waveforms (i.e., P-wave, QRS-complex, and T-wave) [2].

ECG signals are low bioelectrical signals and are very changeable; therefore, some noises may influence the characteristics of the signals, such as baseline wander, motion artifact, and muscle artifact [2, 5]. This makes ECG delineation through visual examination more difficult for cardiologists [6]. Many works in the literature have explored ECG delineation algorithms based on machine learning and digital signal processing [7,8,9,10,11,12,13]. However, a major limitation is that most previous studies used only a single-lead or several leads to generate the delineation algorithm. Single-lead ECG is commonly used for basic heart monitoring, irregular rhythm checks, or observation of the effects of exercise on the ECG [14]. Hence, single-lead ECG monitoring may less accurately measure cardiac electrical activity [15]. In addition, another limitation of using single-lead ECG is a lack of sufficient validation data in actual clinical practice [16]. In our previous study [3, 5, 17], we proposed a state-of-the-art deep learning (DL) algorithm to classify P-wave, QRS-complex, T-wave, and other straight lines or no deflection of ECG (isoelectric line) by using a single-lead, and the study produced outstanding performance (above 98% accuracy and precision). However, the model could not be properly implemented—as specific diagnose, such as myocardial infarction that will show significant ST segment elevation, is mandatory established by examining the number of leads (12-lead ECG) to observe morphological changes, accurate diagnosis and prompt therapeutic measures [18].

In realistic clinical settings, 12-lead ECG, including six limb leads (I, II, III, aVF, aVR, and aVL) and six chest leads (V1, V2, V3, V4, V5, and V6), is a standard test performed in primary and intensive care units and can provide more valuable information than single-lead ECG [19]. The 12-lead ECG is a practical and cost-effective strategic alternative to routine echocardiography [20]. Some heart diseases require a standard 12-lead ECG observation because each ECG signal has a different heart vector orientation.

Automated computer analysis of standard 12-lead ECG has gained significance in the medical diagnosis process and is becoming more prevalent growing [21]. However, the use of a delineation algorithm for 12-lead ECG is still largely unexplored. Conventional algorithms based on wavelet transform have been implemented for P-wave, QRS-complex, and T-wave detection in 12-lead ECG [22]. However, the algorithm lacks feature analysis and has a high degree of uncertainty due to the subjective measurement aspect. Machine learning (ML) approach, which is a subset of DL has also proposed to waveform delineation in 12-lead ECG [23, 24]. Unfortunately, ML typically requires more ongoing human intervention to feature representation. Whereas with ML systems, we need to identify the applied features based on the type of data, and DL system learns the features without additional human intervention. The model of DL uses distinct layers to learn and discover high-level features from the data on its own. In addition, the delineation of 12-lead ECG is challenging because the resultant ECG pattern may vary when the location of the electrodes on the chest wall is varied. Further, the 12-lead ECG delineation may not be able to detect the boundary of ECG waveforms, as each of the 12-lead ECG represents varying morphology and a different direction of cardiac activation in a three-dimension shape.

To enhance the drawbacks, this study aimed to explore an algorithm for delineating 12-lead ECG using an automated feature extraction method—DL. In our previous study [3], the DL has obtained the outstanding results with the performance above 98% accuracy and precision for single-lead ECG beat delineation. Therefore, in the current study, we have improved and adjusted the previous model to 12-lead ECG delineation. The DL architecture consists of a convolutional neural network (CNN) for feature extraction and bidirectional long short-term memory (bidirectional LSTM/BiLSTM) as a classifier for 12-lead ECG. BiLSTM can be learned to use all available input data for a specific timeframe in the past and future. Both techniques can enhance the ECG waveform classification performance to automatically extract features from the input signals in 12-lead ECG. To the best of our knowledge, we are the first to implement and explore automatically high-level feature representation using DL to delineate 12-lead ECG. Our goal was to improve the delineation model with automated feature representation to 12-lead ECG. For this paper, the main contributions were as follows:

  1. 1.

    Proposing a 12-lead ECG delineation model using CNN for feature extraction and BiLSTM as a classifier. The convolution layer of CNN can generate local features of the ECG signal sequence to recognize regional patterns in the convolution window, and BiLSTM appropriates for sequential data processing based on forward and backward time steps;

  2. 2.

    Classifying ECG waveforms (i.e., P-wave, QRS-complex, T-wave, and isoelectric line) in two scenarios using beat-based and patient-based for 200 recordings (patients) with total 14,588 beats;

  3. 3.

    Training, validating and testing the ECG waveforms by lead-to-lead with total of 12-lead ECG using beat-based and patient-based segmentation approach.

Materials and methods

12-lead ECG

The 12-lead ECG is a standard non-invasive test in cardiology clinical practice [25]. The 12-lead ECG represents the recorded electrical activity of the heart from 10 electrodes on the body surface. An electrode is a conductive pad that is attached to the skin to record electrical activity, which is placed on different parts of limb and chest of patient. The 12-lead ECG consists of six limb leads (I, II, III, aVR, aVL, and aVF) and six chest leads (V1, V2, V3, V4, V5, and V6) [26]. Each of the 12 leads graphically describes the electrical activity of the heart. It represents a particular orientation in space (right arm, RA; left arm, LA; left foot, LL). Table 1 lists the 12-lead ECG electrical activity based on the anatomical relations view.

Table 1 The 12-lead ECG electrical activity [27, 28]

Data preparation

The proposed algorithm was evaluated on the Lobachevsky University database (LUDB) of ECG signals, which contained 200 records from 200 subjects (healthy volunteers and patients with various heart abnormalities) [29, 30]. The LUDB had a 10-s 12-lead ECG (I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, and V6) digitized at 500 samples per second. The total of 58,429 annotated waves consisted of 16,797 P-waves, 21,966 QRS-complexes, and 19,666 T-waves. The boundaries and peaks of the ECG waveform were manually annotated by two certified and practicing cardiologists. The sample plot of 12-lead ECG signals is presented in Fig. 1.

Fig. 1
figure 1

Samples of sinus rhythm in 12-lead ECG signals of LUDB

Noise removal

Changes in ECG waveforms indicate a cardiac illness that may occur for any reason. ECG signals are enhanced by eliminating various kinds of noise and artifacts. This study proposed discrete wavelet transform, which is a frequently used denoising technique that offers a valuable option for denoising ECG signals [31, 32]. Some wavelet families for ECG signal, such as symlets (sym), daubechies (db), and bior, were implemented to identify and analyze the type of wavelet that will obtain the best signal denoising result. Increasing the signal-to-noise ratio (SNR) to train the DL model is critical. A higher SNR implies better and more trustworthy ECG data. Based on the highest SNR results, bior wavelet or Bior6.8, was the best wavelet function and was chosen for ECG signal denoising. The SNR value obtained was 8.44 dB.

Segmentation

The ECG waveforms were segmented beat to beat in simple way, the process is cutting the ECG signal based on the window size and sampling frequency. For each segmented time window, it contains one heartbeat and has a length of 512 nodes. Each window size has a start of P-wave1 to the start of P-wave2. If one heartbeat is less than 512 nodes, we perform zero padding technique by adding the value 0 (zero) until the signal has a length of 512 nodes. The LUDB provided the label annotation (ground truth) as waveform onset “(” and offset “)” annotations for P-wave, QRS-complex, and T-wave, respectively. If the annotation doesn’t provide the label onset “(” and offset “)”, the beats segmentation are excluded. A total of 58,429 annotated waves were selected to generate the DL model as ground truth.

Deep learning model

In our previous study [3], we generated the stacked convolutional layer and BiLSTM model for single-lead ECG delineation using the QT database. In the current study, we adjusted the model for the 12-lead ECG signal using LUDB. The convolution layers can extract deep features from ECG signal data points, and BiLSTM with forward and backward schemes help us learn from future and previous representations. We have trained the all beats from the start of P-wave1 to the start of P-wave2 for each lead (lead-by-lead). The rectified linear unit function (ReLU) was adopted with the convolution layer (8, 16, 32, 64, 128, 256, 512, 1024, and 2048 filters). We generated the 13 DL models using the convolution layers, LSTM and BiLSTM. Each fine-tuned hyperparameter of the 13 models is listed in Table 2. We fine-tuned—followed by applying varying convolution layers, from 1 to 4 with the LSTM classifier (Models 1–4) and 1 to 9 with the BiLSTM classifier (Models 5–13). All models were trained over 300 epochs, with a batch size of 8, a learning rate of 10–5, and categorical cross-entropy as the loss metric.

Table 2 Hyperparameter tuning of models

Among the hyperparameter-tuned models, the best model was Model 11. The pseudocode algorithm of the best model can be seen in Algorithm 1. The seven convolution layers and BiLSTM architecture were proposed as the ECG waveform classification model. The model takes the ECG signals as input and output multiclass classification (P-wave, QRS-complex, T-wave, and isoelectric line). The proposed network architecture is represented in Fig. 2. As shown in Fig. 2, the network consisted of seven convolution layers with varying filter numbers (8, 16, 32, 64, 128, 256, and 512 filters) and a kernel size of 3 to extract features. Convolution layers were used to automatically extract features and generate feature maps [33]. Feature maps were convolved by a trained kernel or filter and represent the intensity of ECG waveform features. A waveform with a similar intensity can be classified into the same class. Rectified linear unit layers performed nonlinear activation. The softmax function returns a probability of class membership for each class label, attempting to best approximate the predicted target for each input.

figure a

Algorithm 1. The pseudocode of CNN-BiLSTM

Fig. 2
figure 2

The proposed methodology of the 12-lead ECG delineation model

Results and discussion

For this experiment, we segmented the waveform by beat-to-beat (start of P-wave1 to start of P-wave2) for each lead ECG. For the segmentation of beats, we have experimented based on two approaches: beat-based and patient-based. In this study, both approaches have be done to analyze the consistency of the performance set (training, validation and testing (unseen)) to avoid data leakage problems. The total of 200 records have randomly segmented (beat-to-beat segmentation). The total beat was 14,588 beats. For beat-based segmentation, the total of beats are consisting of 11,666 beats for training, 1,749 beats for validation, and 1,173 beats for testing (refer to Table 3). In addition, for patient-based segmentation, the total of beats are consisting of 13,331 beats for training, 645 beats for validation, and 621 beats for testing (refer to Table 4). We have used patients 1–180 as training set, patients 181–190 as validation set, and patients 191–200 as testing set. We have trained, validated and tested all beats for each lead (lead-by-lead).

Table 3 The total of P-wave1 to start of P-wave2 beat-based segmentation from 12-lead ECG signals
Table 4 The total of P-wave1 to start of P-wave2 patient-based segmentation from 12-lead ECG signals

The testing set was used as an unseen set because it showed model data that had never been seen before. The total number of beats segmentation is different due to only ground truth that provided the annotations waveform onset “(” and offset “)” are included. There are some unannotated waves. Also, some abnormal morphologies changes of ECG waveform, such as only upwards, only downwards, biphasic negative-positive, or biphasic positive-negative have affected the bias of total number of beats segmentation. For ECG waveform classification for 12-lead ECG signal, we used some performance metrics, such as accuracy (ACC), sensitivity (SEN), specificity (SPE), precision (PRE), and F1-Score (F1). As stated before, this study used 13 deep learning models for classification tasks.

Beat-based segmentation

To generate the best model, we firstly experimented a beat-based segmentation approach and implemented the ACC, SEN, SPE, PRE, and F1 for the results of the 13 models (Table 5). Table 5 presents the performance results of 13 models with different hyperparameters. As shown in Table 5, the results of ACC, SEN, SPE, PRE, and F1 became higher as the convolution layers increased. LSTM was implemented in Models 1–4, and the highest ACC and PRE obtained were 98.05% and 93.06%, respectively. However, when the classifier was modified to BiLSTM, the highest results, 98.82% for ACC and 95.94% for PRE, were achieved (Model 11). The results seemed slightly insignificant when the eight and nine convolution layers were added (Models 12 and 13). Therefore, we proposed Model 11 as the best model for 12-lead ECG waveform classification in this study.

Table 5 The performance results of 13 hyperparameters tuning models (beat-based segmentation)

Figure 3 shows the 12-lead ECG performance results of Model 11 (the best model). As shown in Fig. 3a, we tested the best model with validation data (1749 beats) and found that the performance results of each lead showed no significant difference (above 93%). However, the highest precision and accuracy were achieved by the chest lead and lead V3 (99.03% and 96.53%, respectively). Also, lead V3 had good performance in all metrics. The good performance of lead V3 shows a nonspecific sign that a wide variety of ECG abnormalities, such as right bundle branch block due to inverted T-waves, could occur. Commonly, lead V3 is placed diagonally between leads V2 and V4. The ECG morphology of lead V3 observes the anterior wall of the left ventricle and is therefore named the anterior lead.

Fig. 3
figure 3

The performance results of 12-lead ECG from the best model to the validation and testing set (beat-based)

Furthermore, the worst performance result of the 12-lead ECG was lead III. The sensitivity and precision were only 93.84% and 94.78%, respectively. We had a challenge observing lead III, as its morphology was inverted almost the entire length of the ECG recording (refer to Fig. 1). Lead III kept track of the inferior aspect of the left ventricle. Figure 3b shows that model 11 was also tested using testing data (unseen). A total of 1173 beats were tested as unseen data. Like the results of the validation data, the testing data results showed no difference, and the performance result was above 93%. The best performance was also achieved by the chest lead (V3–V5), while the worst was lead III. Since the results of the chest lead outperformed the limb lead, this may reveal posterior ST-segment elevation myocardial infarction (STEMI).

The confusion matrix (CM) has visualized to measure the performance of actual and predicted values (refer to Fig. 4). Each diagonal element of CM represents a successfully classified result. The off diagonals of the CM show the misclassified results. In light of this, the ideal classifier will have a CM with just diagonal members and all other values set to 0. Figure 4 shows the most misclassified occurs in isoelectric line, which falsely classified as P-wave, QRS-complex and T-wave and vice versa. However, among the P-wave and QRS-complex, the T-wave has the highest misclassified in all ECG 12-lead. The detection of T-wave is arduous due to the low amplitude, varying morphology (inverted, upwards, downwards, or biphasic T-wave) and its overlapping with P-wave. The minimal error of T-wave classification occurs mostly in lead I, II, and V2 – V5. T-wave represent the ventricular repolarization, and normal T-waves are upright in those leads.

Fig. 4
figure 4

The confusion matrix of the best model in the ECG 12-lead performance

To present the performance results of each ECG waveform class (i.e., P-wave, QRS-complex, T-wave, and isoelectric line), we used the boxplot (also called a box and whisker plot) to display the distribution of the results based on the minimum, first quartile (Q1/25th percentile), median (the second quartile, Q2/50th percentile), third quartile (Q3/75th percentile), interquartile range (IQR), and maximum. Figure 5 shows that the values in the performance results were spread out. Leads I–III were achieved around a minimum of 86% PRE, and around 91% PRE in leads aVR, aVL, and avF. In chest leads (V1–V6), the performance changes were significant, with a minimum of 95% PRE. All boxplots of each lead do not suspect any anomalies due to errors in data collection. To reduce or remove outliers, the whisker on the appropriate side was drawn to 1.5 IQR rather than the data minimum or maximum.

Fig. 5
figure 5

The boxplot of P-wave, QRS-complex, T-wave, and isoelectric line performance in 12-lead ECG

The result of the best model that shows the ground truth and the proposed CNN-BiLSTM model in testing data (unseen) is presented in Fig. 6. The ground truth is the ECG boundary of LUDB annotation from two cardiologists. Figure 6 shows that the error of delineation tends to be T-wave (yellow color) misclassified. T-wave represents ventricles repolarization and, basically, is difficult to detect, which is why a rather comprehensive discussion is needed. The P-wave (red color) indicates atrial depolarization and is mostly clear in lead II. In addition, QRS-complex (blue color) shows outstanding performance in almost all chest leads.

Fig. 6
figure 6

The comparison ECG waveform classification between ground truth and proposed CNN-BiLSTM model based on testing data (beat-based segmentation)

As seen in Fig. 6, the classification of ECG waveform boundary in 12-lead ECG is a challenging task. The morphological characteristic of each lead affects the delineation performance. Each lead represents the difference in electrical potentials measured at two points in space. Leads I, II, III, aVR, aVL, and aVF provide the 12 perspectives of the activity of the heart through the frontal plane, while the rest of the leads record the voltages of the heart onto the horizontal plane. The degree of morphological changes is determined by the ECG lead, the amount of the shift, the direction of displacement, and the ECG segment chosen for analysis.

Patient-based segmentation

The hyperparameters tuning of Model 11 have retrained and experimented to a patient-based segmentation. With the same hyperparameters tuning, the performance results of patient-based segmentation can be presented in Fig. 7. Figure 7 presented the results of ACC, SEN, SPE, PRE and F1 from patient-based segmentation. Compared to the performance results of beat-based, beat-based segmentation outperformed patient-based segmentation even using the same hyperparameters. The average ACC of 12-lead achieved 93.72% for testing set, smaller than ACC of beat-based segmentation that was achieved 95.89%. Also, the performance results of SEN, SPE, PRE and F1 were decreased. In clinical practice, the characteristics of ECG signals in each patient are different. The varying ECG morphology of each patient can be considered, due to LUDB has varying heart rhythm types with related to heart disorders. The visualization of ECG waveform classification using testing set (unseen) can be presented in Fig. 8. Regardless of beat-based and patient-based segmentation, the performance results are well-performed with the ACC, SEN, SPE, PRE and F1 above 93%. We concluded both models can be implemented for delineation task. However, from the results comparison, the beat-based segmentation approach can be proposed in this study.

Fig. 7
figure 7

The performance results of 12-lead ECG from the best model to the validation and testing set (patient-based)

Fig. 8
figure 8

The comparison ECG waveform classification between ground truth and proposed CNN-BiLSTM model based on testing data (patient-based segmentation)

This investigation compares the proposed CNN-BiLSTM to other recurrent network algorithms, i.e., gated recurrent unit (GRU, bidirectional GRU (BiGRU) and unidirectional LSTM. They have been used as classifiers for ECG waveform classification in 12-lead ECG. The results are presented in Table 6. The varying results of all performance results are not significant; they ranged between 92.41% to 99.21%. However, the proposed BiLSTM classifier outperformed other recurrent network algorithms (GRU, BiGRU, and LSTM) for this investigation. In some cases, LSTM achieved more powerful results if compared to GRU though GRU has a simpler architecture with two gates (update and reset gates). It can be shown the proposed CNN-BiLSTM is worthy of to 12-lead ECG delineation task.

Table 6 The performance results of recurrent network algorithms

The proposed CNN-BiLSTM model also can be compared to other DL architectures as shown in Table 7. Previous studies have implemented a single-lead or multiple-leads to classify ECG waveform (i.e., P-wave, QRS-complex, and T-wave) using LUDB [34,35,36,37]. Some studies proposed U-Net architecture for automatic ECG delineation [34, 36]. Chen et al. [34] proposed U-Net architecture, which the encoder and decoder are being the main components of the proposed architecture. They have segmented the P, QRS, and T-waves using a single lead and achieved the sensitivities 99.88%. Moskalenko et al. [36] have also experimented U-Net architecture, which used two convolution layers and connected sequentially with MaxPooling layers. Their work used a lead II (single lead), and obtained the average of sensitivity and precision for the P, QRS, and T-waves are 99.23% and 98.99%, respectively.

Table 7 Comparison of the results of the delineation model to other DL architectures

Liu et al. [35] have proposed other architecture of CNN, ResNet, and hybrid to LSTM for P, QRS, and T-waves classification in a single lead. They experimented many kinds of ECG database, such as QTDB, LUDB, MITDB and BUT PDB. They used four convolution layers with the varying kernel size (4, 6, and 8). LSTM layer is connected to the residual network for achieving deep features. The proposed hybrid neural network has obtained the average of 99.83% sensitivity. Jimenez-Perez et al. [37] have explored the W-Net architecture, which it is the application of a second U-Net whose input is the output of the first U-Net. They have solved the issue of performance and complexity trade-off by using the efficient channel attention (ECA) mechanism. They experimented the ECG delineation task in single and multi-lead approach. For single lead, the precision of P, QRS, and T waves achieved 99.27%, 99.31% and 98.73%, respectively. However, when they applied the proposed model in multi-lead, the precision decreased to 98.90%, 99.24% and 98.24%, for P, QRS, and T-waves, respectively.

These works showed their proposed ECG delineation model is well-performed, however the scope of experimental study limited to the use of single lead ECG (mostly lead II). Also, in other perspective of the use of CNN in those studies, specifically for the convolution layers are excellent for feature extraction. Convolution layers may take advantage of any existing spatial and temporal patterns in the data.

Therefore, this study implemented the convolutional layers and BiLSTM to classify the 12-lead ECG waveform boundary. Also, it can be a preliminary task to develop and improve ECG delineation performance in cardiology clinical practice. The treatment of ECG signal processing in single and 12-lead ECG is different. The 12-lead ECG delineation is a more challenging task due to varying lead morphology. The repolarization and depolarization of ECG waveform can be arduous to handling. In this study, we have only experimented a single ECG database (LUDB), which has a single frequency sampling (FS). The proposed model has not been generalized to other 12-lead ECG databases due to the availability of data being limited. It can be our limitation for a preliminary task to generate the automatic 12-lead ECG delineation. The varying 12-lead ECG database can be explored for the future, with varying noise-handling technique and FS.

Conclusion

The automatic delineation of the ECG main waveform in 12-lead ECG poses a challenge due to the characteristics of morphology appearance in each lead. By examining changes in ECG signal morphology, cardiologists can observe multiple heart disease processes. This study aimed to explore and improve the delineation model using the DL algorithm to classify the P-wave, QRS-complex, T-wave, and isoelectric line in a standard 12-lead ECG. This study generated the 13 hyperparameter tuning models and identified the best models using convolutional layers and BiLSTM. Also, for beat segmentation, we have experimented a beat-based and patient-based segmentation. The performance results were achieved above 95% ACC, SEN, SPE, PRE, and F1-score for beat-based segmentation. In addition, for patient-based segmentation approach, the performance results were achieved above 93% ACC, SEN, SPE, PRE, and F1-score. The performance results of beat-based segmentation outperformed the patient-based segmentation. Regardless of both performance results, any beat segmentation approach can be considered. The results can be proposed for a preliminary task to 12-lead ECG delineation. ECG examinations might be misinterpreted due to inaccuracies in electrode placement and variances in interindividual human anatomy. In the future, the unsupervised learning approach by training the network to remove ECG noise can be explored to replace the conventional wavelet transform. Also, the standard 12-lead ECG can be generalized for morphological changes diagnosis, not only limited to single-lead observation.

Availability of data and materials

The datasets generated and/or analysed during the current study are available in the PhysioNet:Lobachevsky University Electrocardiography Database repository, https://physionet.org/content/ludb/1.0.1/.

Abbreviations

ACC:

Accuracy

BiGRU:

Bidirectional GRU

BiLSTM:

Bidirectional long short-term memory

CNN:

Convolutional neural network

DL:

Deep learning

ECG:

Electrocardiogram

F1:

F1-Score

GRU:

Gated recurrent unit

LA:

Left arm

LL:

Left foot

LUDB:

Lobachevsky University database

LSTM:

Long short-term memory

ML:

Machine learning

PRE:

Precision

ReLU:

Rectified linear unit function

RA:

Right arm

SEN:

Sensitivity

SNR:

Signal-to-noise ratio

SPE:

Specificity

References

  1. Hammad M, Maher A, Wang K, Jiang F, Amrani M. Detection of abnormal heart conditions based on characteristics of ECG signals. Measurement. 2018;125:634–44.

    Article  Google Scholar 

  2. Wang J, Li R, Li R, Fu B. A knowledge-based deep learning method for ECG signal delineation. Futur Gener Comput Syst. 2020;109:56–66.

    Article  Google Scholar 

  3. Nurmaini S, et al. Beat-to-beat electrocardiogram waveform classification based on a stacked convolutional and bidirectional long short-term memory. IEEE Access. 2021;9:92600–13. https://doi.org/10.1109/ACCESS.2021.3092631.

    Article  Google Scholar 

  4. Chen M, Wang G, Chen H, Ding Z. Adaptive region aggregation network: unsupervised domain adaptation with adversarial training for ECG delineation. In: ICASSP, IEEE Int. Conf. Acoust. Speech Signal Process. - Proc., vol. 2020-May. 2020. p. 1274–1278. https://doi.org/10.1109/ICASSP40776.2020.9053244.

  5. Tutuko B, et al. DAE-ConvBiLSTM: End-to-end learning single-lead electrocardiogram signal for heart abnormalities detection. PLoS One. 2022;17(12):e0277932.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Peimankar A, Puthusserypady S. DENS-ECG: a deep learning approach for ECG signal delineation. Expert Syst Appl. 2021;165:113911. https://doi.org/10.1016/j.eswa.2020.113911.

    Article  Google Scholar 

  7. Lin C, Mailhes C, Tourneret J-Y. P-and T-wave delineation in ECG signals using a Bayesian approach and a partially collapsed Gibbs sampler. IEEE Trans Biomed Eng. 2010;57(12):2840–9.

    Article  PubMed  Google Scholar 

  8. Xia Z, et al. Real-time ECG delineation with randomly selected wavelet transform feature and random walk estimation. In: 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). 2018. p. 1–4.

  9. Fu D, Xia Z, Gao P, Wang H, Lin J, Sun L. ECG delineation with randomly selected wavelet feature and random forest classifier. IEICE Trans Inf Syst. 2018;101(8):2082–91.

    Article  Google Scholar 

  10. Spicher N, Kukuk M. Delineation of electrocardiograms using multiscale parameter estimation. IEEE J Biomed Health Inform. 2020;24(8):2216–29.

    Article  PubMed  Google Scholar 

  11. Arini PD, Liberczuk S, Mendieta JG, Santa María M, Bertrán GC. Electrocardiogram delineation in a Wistar rat experimental model. Comput Math Methods Med. 2018;2018:2185378.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Sharma LD, Sunkaria RK. Detection and delineation of the enigmatic U-wave in an electrocardiogram. Int J Inf Technol. 2021;13:2525–32.

  13. Darmawahyuni A, et al. Delineation of electrocardiogram morphologies by using discrete wavelet transforms. Indones J Electr Eng Comput Sci. 2021;22(1):159–67.

    Google Scholar 

  14. Breen CJ, Kelly GP, Kernohan WG. ECG interpretation skill acquisition: a review of learning, teaching and assessment. J Electrocardiol. 2022;73:125–8.

  15. Kwon S, et al. Validation of adhesive single-lead ECG device compared with holter monitoring among non-atrial fibrillation patients. Sensors. 2021;21(9):3122.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Bayoumy K, et al. Smart wearable devices in cardiovascular care: where we are and how to move forward. Nat Rev Cardiol. 2021;18(8):581–99.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  17. Nurmaini S, Tondas AE, Darmawahyuni A, Rachmatullah MN, Effendi J, Firdaus F, Tutuko B. Electrocardiogram signal classification for automated delineation using bidirectional long short-term memory. Inform Med Unlocked. 2021;22:100507.

  18. Vogiatzis I, et al. The importance of the 15-lead versus 12-lead ECG recordings in the diagnosis and treatment of right ventricle and left ventricle posterior and lateral wall acute myocardial infarctions. Acta Inform Med. 2019;27(1):35.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Zhang D, Yang S, Yuan X, Zhang P. Interpretable deep learning for automatic diagnosis of 12-lead electrocardiogram. Iscience. 2021;24(4):102373.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Kagiyama N, et al. Machine learning assessment of left ventricular diastolic function based on electrocardiographic features. J Am Coll Cardiol. 2020;76(8):930–41.

    Article  PubMed  Google Scholar 

  21. Ribeiro AH, et al. Automatic diagnosis of the 12-lead ECG using a deep neural network. Nat Commun. 2020;11(1):1–9.

    Google Scholar 

  22. Yochum M, Renaud C, Jacquir S. Automatic detection of P, QRS and T patterns in 12 leads ECG signal based on CWT. Biomed Signal Process Control. 2016;25:46–52.

    Article  Google Scholar 

  23. Mehta SS, Lingayat NS. Development of SVM based classification techniques for the delineation of wave components in 12-lead electrocardiogram. Biomed Signal Process Control. 2008;3(4):341–9.

    Article  Google Scholar 

  24. Saini I, Singh D, Khosla A. P-and T-wave delineation in ECG signals using support vector machine. IETE J Res. 2013;59(5):615–23.

    Article  Google Scholar 

  25. Rizzo C, Monitillo F, Iacoviello M. 12-lead electrocardiogram features of arrhythmic risk: a focus on early repolarization. World J Cardiol. 2016;8(8):447.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Goldberger A. Goldberger’s clinical electrocardiography. Philadelphia: Elsevier; 2018. pp. 11–20.

  27. Hampton JR. The ECG made easy e-book. Philadelphia: Elsevier Health Science; 2019.

  28. Meek S, Morris F. ABC of clinical electrocardiography: introduction. I—Leads, rate, rhythm, and cardiac axis. BMJ. 2002;324(7334):415.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Kalyakulina A, et al. Lobachevsky university electrocardiography database (version 1.0. 0). Cambridge: PhysioNet.[Google Sch.]; 2020.

    Google Scholar 

  30. Kalyakulina AI, et al. Ludb: a new open-access validation tool for electrocardiogram delineation algorithms. IEEE Access. 2020;8:186181–90.

    Article  Google Scholar 

  31. Lin H-Y, Liang S-Y, Ho Y-L, Lin Y-H, Ma H-P. Discrete-wavelet-transform-based noise removal and feature extraction for ECG signals. Irbm. 2014;35(6):351–61. https://doi.org/10.1016/j.irbm.2014.10.004.

    Article  Google Scholar 

  32. Aqil M, Jbari A, Bourouhou A. ECG signal denoising by discrete wavelet transform. Int J Online Eng. 2017;13(9). https://doi.org/10.3991/ijoe.v13i09.7159.

  33. Zhang J, Feng W, Yuan T, Wang J, Sangaiah AK. SCSTCF: spatial-channel selection and temporal regularized correlation filters for visual tracking. Appl Soft Comput. 2022;118:108485.

    Article  Google Scholar 

  34. Chen Z, Wang M, Zhang M, Huang W, Gu H, Xu J. Post-processing refined ECG delineation based on 1D-UNet. Biomed Signal Process Control. 2023;79:104106.

    Article  Google Scholar 

  35. Liu J, et al. A novel P-QRS-T wave localization method in ECG signals based on hybrid neural networks. Comput Biol Med. 2022;150:106110.

    Article  PubMed  Google Scholar 

  36. Moskalenko V, Zolotykh N, Osipov G. Deep learning for ECG segmentation. In: Advances in neural computation, machine learning, and cognitive research III: selected papers from the XXI international conference on neuroinformatics, October 7–11, 2019, Dolgoprudny, Moscow Region, Russia. 2020. p. 246–254.

  37. Jimenez-Perez G, Acosta J, Alcaine A, Camara O. Generalizing electrocardiogram delineation: training convolutional neural networks with synthetic data augmentation. arXiv Prepr. arXiv2111.12996. 2021.

Download references

Acknowledgements

We thank the Intelligent System Research Group, Faculty of Computer Science, Universitas Sriwijaya, Indonesia, for supporting this study in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Funding

This research was funded by Intelligent System Research Group, Faculty of Computer Science, Universitas Sriwijaya, Indonesia. The funding body has played role in the design of the study and collection, analysis, interpretation of data, and in writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

AD: Conceptualization, wrote the manuscript, performed the analysis, and methodology. SN: formal analysis, methodology, and funding acquisition. MNR: Designing computer programs, formal Analysis, methodology. PPA: Designing computer programs, data curation. SBPT: Designing computer programs, data curation, and contributed data or analysis tools. AIS: Resources, Visualization Preparation. BT: contributed data or analysis tools and formal analysis. FF: contributed data or analysis tools and formal analysis. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Siti Nurmaini.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Darmawahyuni, A., Nurmaini, S., Rachmatullah, M.N. et al. Improved delineation model of a standard 12-lead electrocardiogram based on a deep learning algorithm. BMC Med Inform Decis Mak 23, 139 (2023). https://doi.org/10.1186/s12911-023-02233-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12911-023-02233-0

Keywords