Fig. 1From: Multi-infusion with integrated multiple pressure sensing allows earlier detection of line occlusionsSingle-pump occlusion detection using the standard deviation (SD). During stable infusion pressure values are assumed to deviate less than 2 times the SD from the mean pressure value. When the pressure exceeds the threshold set at twice the SD, an occlusion is likelyBack to article page