Skip to main content

Table 1 Agreement between BI-RADS classification assessed by visual evaluation and by the Aguida software

From: Design and clinical validation of a software program for automated measurement of mammographic breast density

  

Software

Total

A

B

C

D

BI-RADS

A

143

8

0

0

151

 

B

14

30

1

0

45

 

C

0

5

68

3

76

 

D

0

0

3

16

19

Total

 

157

43

72

19

291