Skip to main content
Fig. 1 | BMC Medical Informatics and Decision Making

Fig. 1

From: Design and clinical validation of a software program for automated measurement of mammographic breast density

Fig. 1

a Clinical validation of the Aguida software for automated measurement of mammographic breast density – Agreement between the software and visual assessment; b Clinical validation of the Aguida software for automated measurement of mammographic breast density – Correlation between the software and visual assessment; c. Clinical validation of the Aguida software for automated measurement of mammographic breast density – Correlation between the difference and the average of visual and software measurements of breast density; d. Clinical validation of the Aguida software for automated measurement of mammographic breast density – Reliability of breast density measurements with the software between two mammograms in the same subjects separated by 1 to 3 years

Back to article page