Table 4 Passing rate of the autoverification system of single tests and whole samples

From: Design and evaluation of a LIS-based autoverification system for coagulation assays in a core clinical laboratory

  n Autoverification results PT APTT TT FBG Whole sample
July 2017 13,416 Pass 12,485 12,389 12,536 12,557 10,453
Fail 931 1027 880 859 2963
Passing rate (%) 93.06 92.34 93.44 93.60 77.91
August 2017 14,477 Pass 13,349 13,413 13,723 13,573 11,354
Fail 1128 1064 754 904 3123
Passing rate (%) 92.21 92.65 94.79 93.76 78.43
September 2017 13,673 Pass 12,784 12,456 12,585 12,932 10,740
Fail 889 1217 1088 741 2933
Passing rate (%) 93.50 91.10 92.04 94.58 78.55
October 2017 13,004 Pass 12,235 12,059 11,989 12,304 10,289
Fail 769 945 1015 700 2715
Passing rate (%) 94.09 92.73 92.19 94.62 79.12
November 2017 14,866 Pass 13,832 13,905 14,203 13,857 11,934
Fail 1034 961 663 1009 2932
Passing rate (%) 93.04 93.54 95.54 93.21 80.28
December 2017 14,263 Pass 12,995 13,378 13,254 13,157 11,258
Fail 1268 885 1009 1106 3005
Passing rate (%) 91.11 93.80 92.93 92.25 78.93
  1. PT Prothrombin time, APTT Activated partial thromboplastin time, TT Thrombin time, FBG Fibrinogen