From: Design and evaluation of a LIS-based autoverification system for coagulation assays in a core clinical laboratory
May 2017
June 2017
n
13,949
14,295
Pass
10,828
11,257
Fail
3121
3038
True-positive
858
892
True-negative
10,823
False-positive
2263
2146
False-negative
5
0
Sensitivity
99.42%
100%
Specificity
82.71%
83.99%
Passing rate
77.63%
78.75%