Table 2 Clinical reference intervals, critical range, 5 and 95% percentiles, and the limited range

From: Design and evaluation of a LIS-based autoverification system for coagulation assays in a core clinical laboratory

 

Clinical reference intervals

Critical value

Percentiles (5 and 95%)

Limited Range

PT

11.00–14.30 s

≤ 9.00 s or  ≥  70.00 s

11.80–16.30 s

11.00–16.30 s

APTT

32.00–43.00 s

≤ 15.00 s or  ≥  100.00 s

30.40–46.40 s

30.40–46.40 s

TT

14.0 0–21.00 s

> 150.00 s 

15.00–19.40 s

14.00–21.00 s

FBG

2.00–4.00 g/L

< 1.00 g/L

2.21–6.51 s

2.00–6.51 g/L